当前位置: 首页 > 康体医疗 > 急救呼吸机 > BLS型急救呼吸机


BLS型急救呼吸机

BLS型急救呼吸机
BLS型急救呼吸机
 
 
产品简述:
技术参数 CAREvent®BLS
潮气量范围 0.2-1.1L
每分钟呼吸频率 20-12
精确档位调节(呼吸频率BPM/潮气量ML) 20/200,15/300,15/400,12/600,12/800,12/1100
I:E比率 1:2
自动流量范围 12-39.6LPM
手动流量范围 依照自动档设置
同步呼吸功能 无
输入压力范围 45PSI(3.1kg)-70PSI(4.8kg)
最大气道压力声音报警 60cmH2O,可选45 cmH2O
操作温度 -18℃ 至50℃0℉至122℉
存放温度 -40℃ 至60℃-40℉至140℉
气瓶气体使用时间("D"型铝制满瓶415升) 20BPM/0.2L 100分钟(大约)12BPM/1.1L 30分钟(大约)
产品说明:
       加拿大O-Two Systems生产的吉斯系列急救呼吸机可以为呼吸衰竭、心肌梗塞或、呼吸停止或缺氧患者提供有效、安全的人工呼吸手段,轻小、便携,经久耐用,按照急救医疗和紧急复苏的要求,是很好的急救呼吸机,也可以在任何地方使用。可连接到氧气瓶、墙上的管道气源、空气源上使用。
  这款急救呼吸机设计为气动、定时、定容呼吸机,提供气动呼吸机、吸引器、面罩供氧多种功能。手动控制可以在自动设置状态下介入,允许使用者按期望的流量、潮气量手动控制。所有控制部分都集中在呼吸机前部朝向操作者的一面,呼吸机包括手动控制按钮,这种设计可使操作者的双手始终压在面罩上,以保证在呼吸机工作时有很好的脸罩密封性。气动回路可以由氧气或压缩空气驱动,该机连上气源和压力表即可马上投入使用。在有氧气源急救的情况下,呼吸机在急救复苏过程可提供100%的纯氧,符合美国心脏协会(JAMA)的要求,可提供生理标准的呼吸频率和潮气量。遵循JAMA的要求,设置具有气道压力限制并可听报警。此急救呼吸机尤其适用于儿童和成人的急救复苏、院前抢救、急救病人转运,院内科室之间急救转运。

 

 